|
|
|
Pisa 2002 |
|
| Page 1 of 1 |
At present I have just these 7 papers. I hope to upload the free papers soon.
Harper
Meakin
Franks
Leier
Rolly
Versichelen
White
Please note:
These abstracts were scanned from the set circulated to delegates. I have not sought to reproduce all figures (though only a few are missing). I have left the text largely unedited, hence variations in spelling and representation of decimals. I have dared to correct a few obvious grammatical errors. Please advise me of any crass mistakes. The formatting will be sorted soon.
IS OXYGEN AN ACCEPTABLE CARRIER GAS?
Dr. Nigel J.N. Harper
Manchester
There are three pragmatic choices for the carrier gas used to deliver inhalational agents during anaesthesia. The same choice applies, of course, during total intravenous anaesthesia. The use of nitrous oxide during anaesthesia is declining worldwide for reasons that are discussed separately. From the specific viewpoint of low flow anaesthesia, a significant disadvantage of the use of nitrous oxide is the need to use an initial high flow of fresh gas to accommodate the high uptake during the first 5-10 minutes. If oxygen alone is used, the minimum theoretical gas flow approximates to the consumption of oxygen by the patient, ie. 250-350 ml/min. The only constraint in this case is the requirement to supply sufficient volatile agent to the lungs, because the uptake of volatile agent is also high at this time. Mapleson (1998), in a theoretical paper, suggested that the initial fresh gas flow of pure oxygen should approximate to the minute volume, but many would wish to use lower flows from the outset. Several solutions to this problem present themselves. The first is to place the vaporizer inside the breathing circuit. The second is to inject the volatile agent in a Quantitative fashion directly into the breathing circuit. Many anaesthetists have taken the step of replacing nitrous oxide with air as the carrier gas. Although this approach has attractions, there are several difficulties. Most current anaesthetic machines do not have a flowmeter for air that is calibrated in sufficiently fine graduations for low flows to be a possibility with any degree of accuracy. A second, unavoidable problem is the f ixed ratio of oxygen to nitrogen, to which oxygen has to be added in a varying proportion. The use of medical air necessitates the installation of an additional pipeline in some cases. In the majority of hospitals, medical air is made on-site by a large compressor. There are several instances where poor maintenance of the compressed air plant has resulted in contamination by oil and other substances.The use of pure oxygen as the carrier gas has many attractions. Safety is , enhanced because it is impossible to deliver a hypoxic mixture. If pure oxygen is used, it is possible to use low flows from the start of anaesthesia. However, before embarking on a major change in anaesthetic practice, it is necessary to be reassured that the potential adverse effects of breathing oxygen are insignificant in practice. These are: pulmonary atelectasis; pulmonary toxicity; awareness during anaesthesia and ventilator/ depression. Toxicity to the eye (retrolental fibroplasia) is well described in neonates but will not be considered here. Volunteers breathing pure oxygen will experience symptoms after a period of time. A vital capacity inspiration is associated with substernal pain, especially in the older subject in whom there is a greater area of the lung with a low ratio of ventilation to perfusion. However, exposure of normal volunteers to pure oxygen for many hours is not associated with significant sequelae. Only when the ambient pressure of oxygen is increased to two atmospheres or above, will central nervous system toxicity be exhibited in the form of "oxygen convulsions".
The exact origin of the concept of nitrogen as the "splint" or "skeleton" of the lung is lost in history. It has long been assumed that a certain partial pressure of nitrogen is required in the alveoli to prevent their collapse and the generation of atelectasis. Once small airways have collapsed, it is clear why alveoli distal to the obstruction become collapsed as a result of the uptake of oxygen into pulmonary capillaries along its partial pressure gradient. It is less clear why greater than usual alveolar collapse should occur de novo. Computerised tomography (CT) provides a sensitive method for detecting atelectasis. Akca and colleagues (1999) raised significant doubts about the role of nitrogen in the avoidance of atelectasis during anaesthesia. In a randomised study, the lungs of patients undergoing colonic resection were ventilated with either 30% or 80% oxygen in nitrogen during surgery and the same gas mixture was administered for two hours in the recovery room via a facemask. There was no significant difference between the groups in the incidence of CT-proven evidence of atelectasis (2.5% v 3,0%). The chosen oxygen concentration of 80% coincides with that observed in the clinical practice of low flow anaesthesia using oxygen as the carrier gas. When a flow of 250-350 ml/min oxygen is supplied to the circle, the inspired concentration of oxygen initially rises towards 100% before falling to approximatelv 80-85% as the result of the elution of nitrogen from body stores into the breathing system.
Pulmonary oxygen toxicity is a real phenomenon and can be readily demonstrated in patients and animal models in the presence of acute lung injury. In patients with acute lung injury, mechanical ventilation with pure oxygen increased the intrapulmonary shunt from 16% to 23% after one hour (Santos, 2000). The central mechanism is the generation of free oxygen-derived radicals. The archetype is the superoxide anion, or superoxide free radical which results from the acceptance of an additional electron into the outer 2P shell of one oxygen atom of the oxygen pair. The unpaired electron confers extreme reactivity to the superoxide free radical which attacks lipids (peroxidation), nucleotides (chromosomal disruption) and S-H containing proteins (many enzymes), A chain of reactions generates further free radicals, including the pernicious hydroxyl free radical, via hydrogen peroxide. Inhalational agents influence the generation of free radicals in a dose-dependent fashion. High concentrations appear to reduce the superoxide-releasing activity of human neutrophils. In anaesthesia-analogous concentrations, isoflurane appears to increase the generation of superoxide free radicals whilst enflurane causes little change and halothane diminishes production (Nakagawara, 1986), It has been suggested that isoflurane preconditions the myocardium against ischaemic insults via the release of free radicals (Mullenheim et al., 2002). The clinical relevance of these important observations is unknown.
Evidence suggests that exposure of patients with normal lungs to pure oxygen does not result in significant lung injury, even after many hours (Singer 1970, Barber 1970). Experiments in rabbits suggest that, after 36 hours, lung water is greater, and alveolar albumin, neutrophils and pro-inflammatory cytokines are increased, but there are no changes in oxygen transfer or lung mechanics (Takao, 1996).
The application of high inspired-concentrations of oxygen appears to have useful antimicrobial effects. In a dramatic study of 500 patients undergoing bowel surgery, Greif et al (2000) demonstrated that postoperative wound infection could be reduced by half (5.2% v 11.2%) simply by increasing the inspired oxygen concentration from 30% to 80% during surgery and for 2 hours in the recovery room. Arterial, subcutaneous and intramuscular oxygen tensions were measured using needle oxygen electrodes. In the group receiving 80% oxygen, the subcutaneous P02 was increased from 7.9 kPa to 14.5 kPa during surgery, suggesting that elevation of local tissue P02 confers some protection against colonisation with bacteria.
It is probable that hyperoxia also has beneficial local antimicrobial effects in the lung. The bactericidal and phagocytic activity of alveolar macrophages normally falls during anaesthesia. Kotani at al (2000) demonstrated that the decline in function in alveolar macrophages harvested from postoperative patients by broncheoalveolar lavage was halved in a group receiving pure oxygen compared with a group who received 30% oxygen. Expression of pro-inflammatory cytokines was 20 times greater in the 100% oxygen group, suggesting potential mechanisms for the apparently beneficial effects of hyperoxia.
It is possible to conclude that the use of oxygen as the carrier gas during anaesthesia is not contraindicated in patients with normal lungs. The use of pure oxygen simplifies low flow anaesthesia in practice. It is probable that hyperoxia confers some protection against bacterial infections. Further work is needed to identify the place of hyperoxic anaesthesia in clinical practice.
REFERENCES
Mapleson WW. The theoretical ideal fresh -gas flow sequence at the start of low-flow anaesthesia. Anaesthesia 1998; 53: 264-272.
Ackra 0 et al. Comparable postoperative pulmonary atelectasis in patients given 30% or 80% oxygen during and 2 hours after colon resection. Anesthesiology 1999;91: 991 -998.
Santos C et al. Pulmonary gas exchange response to oxygen breathing in acute lung injury. Am J Resp Crit Care Med 2000; 161: 26-31.
Nakagawara M et al. Inhibition of superoxide production and Ca2+ mobilization in human neutrophils by halothane, enflurane and isoflurane. Anesthesiology 1986; 64: 4-12.
Do we still need non-rebreathing systems?
Dr G Meakin, Manchester
Breathing systems may be classified broadly into those which are fitted with a means of absorbing C02 (absorber systems - also called rebreathing systems) and those, which do not contain such units (non-absorber - non-rebreathing systems). (1) In the past, concerns about resistance to breathing and apparatus deadspace with the use of absorber systems led paediatric anaesthetists to use mainly non-absorber breathing systems. (2, 3) However, recent concerns for economy and environmental pollution have led to a renewed interest in the use of the circle absorber system in paediatric anaesthesia. (4) In this talk I am going to try to answer the question: - do we still need non-rebreathing systems in paediatric anaesthesia?
Characteristics of the ideal breathing system
The characteristics of the ideal breathing system would include:
• Simplicity
• Lightness
• Ease of use
• Low resistance
• Minimal deadspace
• Low compression volume
• Precise control of the delivered oxygen and anaesthetic concentrations
• Heat and moisture conservation
• Economical use of anaesthetic gases and the ability to scavenge waste gases
Clearly, no single system fulfils all of these requirements; therefore, the choice of breathing system for an individual patient will depend upon clinical circumstances and the patient's condition.
Clinical circumstances
During induction of anaesthesia, we may be more concerned about the ease of use of the breathing system, the ability to control anaesthetic concentrations and a low compression volume than with economy, since high flowrates are going to be required whatever system we choose. During maintenance of anaesthesia, when the inspired and expired anaesthetic concentrations have equilibrated, we may be more interested in using low flow rates to economize the use of anaesthetic agents. However, there may be little scope for low flow methods if the procedure is short. During recovery from anaesthesia economy is not a consideration and we may wish to have the convenience of a T-piece system.
Patients condition
The main patient condition relevant to our question is age. And here we have to be aware that the respiratory system of an infant is disadvantaged in various ways compared with that of an adult. (4) Accordingly, we will want to use a system with minimal resistance and deadspace. We will also want to use a system that is convenient to use in small children and has a low compression volume enabling us to control ventilation adequately in patients with a small tidal volume and low lung compliance. (5)
The circle system consists of a fresh gas inflow, inspiratory and expiratory breathing tubes with unidirectional valves, a spill valve, a reservoir bag, a carbon dioxide absorption unit and a fresh gas inflow. Putting the fresh gas inflow downstream and the spill valve upstream of the absorber improves the efficiency of the system by allowing expired gas from the patient to be vented preferentially.
The main advantages of the circle system are reduced loss of heat and moisture, economical use of anaesthetic gases and reduced operating theatre pollution. (6) From the point of view of their use in paediatric anaesthesia, there have been major concerns about increased resistance due to the inspiratory and expiratory valves and increased apparatus deadspace. (2, 3) Some authors hold the view that the system is relatively complex, bulky and prone to incorrect assembly. (7) It also has a relatively high compression volume and precise control of the delivered oxygen and anaesthetic concentrations can be difficult. (8)
Resistance to breathing during anaesthesia occurs in the breathing system and in the tracheal tube. Traditionally, it is measured in terms of the pressure drop across the equipment at a given flow rate. A study by Orkin and colleagues revealed that in a typical circle system the tubes and the absorber have about equal resistance and together account for about a third of the total resistance of the system (fig.l). (9) Three sets of valves tested had practically the same resistance and accounted for two-thirds of the total resistance. Their data indicate that for an average adult, whose peak flow under anaesthesia is about 35 litre min"', the pressure drop across the complete system should be less than 0.75 cm l-hO while that across the valves should be less than 0.5 cm HzO. Contrary to a widely held belief that the resistance imposed by older anaesthetic breathing systems was unduly high, these values appear to be quite acceptable. (10, 11) For an infant of 9 months, whose peak flow is about 10 litre min' ', the pressure drop across the systems tested by Orkin and colleagues should be less than 0.25 cm l-hO. By contrast, the pressure drop across a 3.5 mm tracheal tube in a 3 month old infant with a peak flow about 6 litre min'1 should be approximately 2.5 cm H2O. (12) These figures suggest that the resistance of the tracheal tube in a young infant is at least 10 times that of the circle system.
Anaesthetised infants cope remarkably well with acute increases in airway resistance as shown by Graff and colleagues. (13) After a moderate increase in airway resistance in 10 anaesthetised infants, there was an immediate increase in the force of breathing, as reflected by oesophageal pressure, so that tidal and minute volumes were maintained for the duration of the test (10 min). The speed of the response suggested a reflex mediated by muscle spindles in the diaphragm. However, the authors also noted that ventilation was maintained at the cost of a three-fold increase in the work of breathing, which could lead eventually to hypercapnia and acidosis as a result of muscle fatigue.Figl.

Litres per minute
Apparatus Deadspace.
Charlton and others have investigated the response of paediatric patients to an increase in apparatus deadspace. (14) These authors found that increasing the deadspace produced an immediate increase in end-tidal CO; in anaesthetised infants and children. However, tidal and minute volumes increased by 40-50% over the next 10 mm so that end-tidal CC>2 tensions returned to baseline values. They concluded that the short term ventilator/ response to an increased deadspace was adequate;
nevertheless, apparatus deadspace should be minimised in equipment designed for children and controlled ventilation should be used liberally in infants.
Compression volume
When we use a breathing system with controlled ventilation (manual or controlled) a certain amount of the tidal volume delivered to the system will be wasted due to compression of gas within the system. This volume, the compression volume, may be calculated from the volume of the breathing system and the ventilation pressure as shown below. (15)
If the volume of circle system is 5000 ml and we apply a ventilating pressure of 20 cm H20, this will increase the ambient pressure (approx 1000 cm HzO) by approximately 2%. Therefore, according to Boyle's Law, 2% x 5000 = 100 ml of gas will be compressed and this is wasted ventilation (compression volume = 100 ml;
compression factor = 100/20 = 5 ml/cm H;>0). If we wish to deliver a tidal volume of 500 ml to an adult patient at a pressure of 20 cm HzO, a ventilator volume of 600 ml will be necessary. In a child with a tidal volume 250 ml, a ventilator volume of 350 ml will be required and in a neonate with a tidal volume of 30 ml the ventilator volume should be 130 ml. Thus, the smaller the patient the greater the proportion of ventilation will be wasted by compression of gas in the system. Furthermore, the use of a standard "adult" circle system will result in a compression volume, which is many times greater than the tidal volume of an infant. The consequences of this are:
• Circle systems give a poor "feel" of the lung compliance in infants. (16)
• In the event of a decrease in lung compliance in an infant being ventilated with an circle system (e.g. spontaneous reversal of muscle relaxants) there may be a substantial reduction in tidal volume being proportional to the compression volume. (17)
• Increasing the inspiratory pressure may be of limited use in this situation since the compression volume will be further increased. (17)
• It may be impossible to ventilate the patient adequately with the circle system.
Jackson Rees system
The original Ayre's T-piece system consisted of a light metal T-tube with a main lumen of 1 cm diameter, and a smaller side tube at right angles to the main lumen through which the anaesthetic gas mixture was introduced: a length of rubber tubing attached to the open end of the T-piece acted as a reservoir for anaesthetic gases. (18) In 1950, Jackson Rees modified the system by attaching an open tailed bag to the reservoir tube, in order to facilitate controlled ventilation. (19) As there are no valves in the system, resistance to breathing is minimal. Apparatus deadspace is also minimal, but as there is no CC>2 absorber in the system fresh gas requirement is high (1.5-2 x minute volume during spontaneous ventilation).
When the T-piece is compared to our "ideal system", we see that it performs quite well in terms of simplicity, lightness, ease of use, low resistance, minimal deadspace, low compression volume and precise control of the delivered oxygen and anaesthetic concentrations. However, as might be expect, it performs badly on heat and moisture conservation, economical use of anaesthetic gases and the ability to scavenge waste gases.
Compression volume
The low volume of the T-piece system is reflected in a low compression volume. (8) Thus, a T-piece with a volume of 70 ml (tubing) plus 500 ml (bag) with a ventilation pressure of 20 cm HzO results in a compression volume of 2% x 570 = 11.4 ml (or less if the bag is only partly full). Therefore, in a neonate with a tidal volume of 30 ml, it will only be necessary to set the ventilator to deliver 41 ml to compensate for the compression volume and the compression volume of the T-piece (11 ml) is approximately 1/10th that of the circle (100 ml). The consequences of this are:
• The delivered tidal volume relates quite closely to the actual tidal volume of the patient, so the anaesthetist should get a good "feel" for the lung compliance.
• If the compliance of the lung decreases (e.g. spontaneous reversal of muscle relaxants), the decrease in tidal volume will also be relatively small being proportional to the compression volume.
• This may be of vital importance when faced with an infant who has developed partial airway obstruction (bronchospasm or laryngospspasm) during induction or recovery from anaesthesia.
Conclusions
While standard circle systems have undoubted advantages for maintenance of anaesthesia they must be used with caution in infants owing to their large compression volumes. (8) This problem is only partly addressed by the use of small bore (15 mm) breathing hoses and a small (800-1000 ml) reservoir bag. In the case of a decrease in lung compliance in a patient with a low tidal volume, it may be difficult to maintain ventilation with a circle system and a T-piece should always be readily available. The T-piece system may still be the system of choice during induction and recovery of anaesthesia and as the sole breathing system for procedures of short duration. In these circumstances the T-piece may be safer and more convenient than the circle system whose main benefits will be less apparent owing to the need for high fresh gas flows. The inability to scavenge waste gases remains a major problem with the T-piece system.
References
1. Conway CM. Anaesthetic breathing systems. BrJAnaesth 1985; 57: 649-657.
2. Adriani J, Griggs T. Rebreathing in pediatric anesthesia: recommendations and descriptions of improvements in apparatus. AnesthesioJogy 1953; 14: 337-347.
3. Stephen CR, Slater HM Agents and techniques employed in pediatnc anesthesia. Anesth Analg 1950; 29: 254-262.
4. Meakin G. Low flow anaesthesia in infants and children. Br JAnaesth 1999;
83: 50-57.
5. Nunn JF. Appliedrespiratory physiology. 3rd ed. London: Butterworths, 1987.
6. Waters RM. Clinical scope and utility of carbon dioxide filtration in inhalation anesthesia. Anesth Analg 1924; 3: 20-22.
7. Hughes DG. Paediatric anaesthetic equipment. In: Mather SJ, Hughes DG, eds. Handbook ofpaedialric anaesthesia. Oxford: Oxford University Press, 1991: 49.
8. Cote CJ; Petkau AJ, Ryan JF, et al. Wasted ventilation measured in vitro with eight anesthetic circuits with and without inline humidification. Anesthesiology 1983;
59: 442-446.
9. Orkin LR, Siegal M, Rovenstein EA. Resistance to breathing by apparatus used in anesthesia. II. Valves and machines. Anesth Analg 1957; 36(2): 19-26.
10. Nunn JF, Exi-Ashi TI. The respiratory effects of resistance to breathing in anesthetized man. Anesthesiology 1961; 22: 174-185.
11. Young TM. Carbon dioxide absorber. Anaesthesia 1971; 26: 78-79.
12. Brown ES, Hustead RF. Resistance ofpediatric breathing systems. Anesth Analg 1969; 48: 842-849.
13. Graft TD, Sewall K, Lim HS, el al. The ventilator/ response of infants of airway resistance. Anesthesiology 1966; 27: 168-175.
14. Charlton AJ, Lindahl SGE, Hatch DJ. Ventilatory responses of children to changes in deadspace volume. BrJAnaesth 1985; 57: 562-568.
15. Fisher DM. Anesthesia equipment for pediatrics. In: Gregory GA, ed. Pediatric Anesthesia, 4th edn. New York: Churchill Livingstone, 2002: 191-216.
16. Carden E, Nelson D. A new and highly efficient circuit for paediatric anaesthesia. Can Anaesth Soc J 1972; 19: 572-582.
17. Stayer SA, Bent ST, Campos CJ, et al. Comparison of the NAD 600 and Servo 900C ventilators in an infant lung model. Anesth Analg 2000; 90; 315-321.
18. Ayre P. The T-piece technique. BrJAnaesth 1956; 28: 520-523.
19. Jackson Rees G. Anaesthesia in the newbom. BMJ 1950; 2: 1419-1422.
Mechanisms of action of inhalational anaesthetics
N. P. Franks
Imperial College of Science, Technology and Medicine
The first anaesthetics to be used in surgery were inhalational agents (nitrous oxide, ether and chloroform), and the anaesthetic properties of the “inert” gas xenon has been known for over half a century. Perhaps surprisingly, however, their mechanisms of action are still uncertain. Indeed, until very recently, there had been no plausible molecular explanation offered for the anaesthetic action of nitrous oxide. Because the potencies of most simple anaesthetics can be accurately predicted by fat solubility (the Meyer-Overton correlation), they have long been considered to be archetypal “non-specific” drugs. This apparent lack of specificity of general anaesthetics is unusual among drugs and simple biophysical correlations caused early workers to suppose that anaesthetics act directly on lipids (the Meyer-Overton Hypothesis). This hypothesis was later extended by postulating that anaesthetic-induced changes in lipid structure and dynamics produce general anaesthesia by disrupting the activity of critical proteins such as membrane ion channels. This traditional view, however, looks increasingly untenable. Quantitative studies over the past 20 years have shown that, at surgical levels, anaesthetic effects on lipid bilayers are extremely small. Moreover, certain lipid-free proteins have been shown to be directly affected by a wide range of anaesthetics, with potencies essentially identical to those for general anaesthesia and crystallographic studies have revealed specific binding sites. Finally, studies with optical isomers of general anaesthetics have revealed stereo-selective effects both on whole animals and on anaesthetic-sensitive neuronal ion channels. Taken together, these findings indicate that general anaesthetics produce their effects at surgical levels by binding directly to proteins rather than to lipids.
Although general anaesthetics are non-specific in the sense that they come in diverse shapes and sizes with no common chemical groupings, their actions on neurons and their ion channels can be surprisingly selective. For example, it has long been known that, in general, synaptic transmission is more susceptible to anaesthetic block than is conduction along axons. Consistent with this observation, the voltage-gated Na+ and K+ channels involved in axonal conductance appear to be insensitive to general anaesthetics. On the other hand, members of a genetically related superfamily of fast neurotransmitter-gated synaptic receptor channels (which include neuronal nicotinic ACh, GABAA, 5‑HT3, and glycine receptors) can be very sensitive to anaesthetics. In some cases, ion channels are affected by inhalational anaesthetics at concentrations well below their minimum alveolar concentrations. The mechanisms involved are beginning to be understood. There is growing evidence that even the simplest of anaesthetic gases, such as xenon, may act at specific targets in the central nervous system. While the GABAA receptor has been identified as the most important single target for intravenous anaesthetics, the relevant targets for inhalational anaesthetics are much less certain and may include the GABAA receptor, certain anaesthetic-activated potassium channels and the NMDA subtype of glutamate receptors.
Many questions remain regarding the molecular and cellular actions of inhalational general anaesthetics. What are the molecular forces that govern the interactions of these agents with their target sites? Are a few critical targets involved, or are there many? What happens at the molecular and cellular levels when inhalational anaesthetics interact with their targets? In this lecture I will address these questions and review the experimental evidence from X-ray crystallography, enzymology and electrophysiology studies concerning the nature and identity of anaesthetic binding sites in the central nervous system.
The Gas-Tight Airway
M Leier
Nuremburg
We are using highly developed anesthesia machines and monitors. The connection between anesthesia machine and patient is still variable with special advantages and disadvantages. The most adequate technique depends on type and length of surgery and, more important, on skill of the anesthetist. Endotracheal intubation is the golden standard, but mechanical ventilation can also be delivered by using a face mask. Between these techniques there is a large group of supraglottic airways. As an alternative for endotracheal intubation or face mask ventilation especially the laryngeal mask airway has reached importance.
An artificial airway should provide a good sealing between respirator and the patients trachea. On the one hand to ensure a gas tight seal, on the other hand to prevent leakage of fluid into the airway. Although endotracheal intubation is the way of choice to secure the airway and to avoid aspiration, there remains some doubt whether a endotracheal tube is always the best way to secure the airway. For example tonsillectomy in children with uncuffed tubes or patients in intensive care units with gastric reflux.
Asai could demonstrate in a benchtop model with a ventilated artificial trachea, that there is a rapid aspiration from above the cuff, even if the cuff is blocked correctly (1). After one hour of cuffed tracheal intubation the mucociliary clearence is a lot more impaired than after laryngeal mask intubation. The risk of retention of secretions by blocking the trachea with a cuff was significantly higher. The disturbance of tracheal mucous velocity caused by intubation reduces the physiologic tracheal cleaning function and can lead to atelectasis and pulmonal infection (2). Side effects and complications of endotracheal intubation may be serious or just unpleasant like coughing, sore throat, hoarseness or swallowing. Complaints are more often after endotracheal intubation than after use of laryngeal mask airway (3).
Supraglottic airways can be divided in three groups.
1. Without protection against aspiration
2. Antiaspiration strategy only by occlusion of the upper esophageal sphincter
3. Antiaspiration strategy by occlusion of the upper esophageal sphincter and drainage of the stomach
The oropharyngeal leak pressure correlates with the suitability of the supraglottic airway for controlled ventilation. There are several methods of testing, to determine the leak pressure. To avoid gastric insufflation and distension a qualitative assessment of gastroesophageal insufflation is advisable. Both, the leak at the neck and gastric insufflation, produce a dSifference between inspiratory and exspiratory volumes (leak fraction), which is easy to measure. A fresh gas flow close to the metabolic rate is the proof for a gas-tight airway. If a fresh gas flow of more than 300 ml/min is needed, you always have to reassure absence of gastric insufflation.
A gas-tight airway is essential for spontaneous breathing as well as for controlled ventilation, because sometimes a change in the planned surgical procedure requires conversion to mechanical ventilation and the use of relaxants.
Surgical procedures with peak airway pressures of 20 mbar or more (e.g. laparoscopy) require a seal pressure of at least 25 mbar to have a minimum safety gap between sealing and airway pressure. Otherwise the device has to be corrected or changed. Actually we investigate the possibility, safety and patients well-being of the LMA Pro-Seal vs. endotracheal intubation during laparoscopic cholecystectomy.
There is no extraordinary risk of pulmonal aspiration reported in literature using a supraglottic airway device. The incidence of postoperative complaints between the supraglottic devices differs, but is in general less than endotracheal intubation (4). Specially children profit by a low rate of coughing, swallowing and laryngospasm.
The gas-tight airway has a strong dependency with patients situation (5). Airway resistance and pulmonal compliance should be normal. The choice of anesthesia management, ventilation procedure, patients positioning and the use of muscle relaxant have a strong influence on the interaction of patient and anesthesia machine. It is obvious, that a supraglottic airway needs an open aperture of glottis. Anesthesia methods should provide a strong depression of airway reflexes. Using drugs like propofol, remifentanyl, alfentanyl and sevoflurane, a muscle relaxant is never required for insertion of the supraglottic airway and may be dispensable, even during smaller abdominal surgery and laparoscopy.
On the one hand, we establish a regime with minimal fresh gas flow to provide a save ventilation without gastric insufflation (6). On the other hand the supraglottic airways have a potential hazard of mucosal damage, because intracuff pressures are 60 mbar and more. Therefore the intracuff pressure should be reduced to Just seal pressure", which may lead to shortage of fresh gas in the anesthesia circle system, if airway conditions are changing.
A good insertion technique, an adapted anesthesia drug regimen and a adequate proficiency is advisable to use supraglottic airways in a optimum range (7). The optimal use is a fresh gas flow close to the metabolic rate and a cuff pressure as low as possible.
In our hospital supraglottic airways, especially LMA-ProSeal are standard for all patients with an empty stomach during short and long lasting surgical procedures, even during laparoscopy and abdominal surgery with and without muscle relaxants. A gas-tight supraglottic airway is a safe and probably a beneficial alternative to endotracheal intubation.
Literature;
1. Asai T, Shingu K, Leakage around high-volume, low-pressure cuffs apparatus A comparison of four tracheal tubes, Anaesthesia 2001 Jan; 56(1);38-42
2. Keller C, Brimacombe J, Bronchial mucus transport velocity in paralyzed anesthetized patients: a comparison of the laryngeal mask airway and cuffed trachea! tube, Anesth Analg. 1998 Jun;86(6): 1280-2
3. Honemann CW, Hahnenkamp K, MollhoffT, Baum JA, Minimal-flow
anaesthesia with controlled ventilation: comparison between laryngeal mask airway and endotracheal tube, EurJ Anaesthesiol. 2001 Jul;18(7):458-66.
4. Oczenski W, Krenn H, Dahaba AA, Binder M, EI-Schahawi-Kienzl I, Kohout S, Schwarz S, Fitzgerald RD. Complications following the use of the Combitube, tracheal tube and laryngeal mask airway. Anaesthesia. 1999 Dec;54(12):1161-5
5. Brimacombe JR, Keller C, Gunkel AR, Puhringer F, The influence of the
tonsillar gag on efficacy of seal, anatomic position, airway patency, and airway protection with the flexible laryngeal mask airway: a randomized, cross-over study of fresh adult cadavers, Anesth Analg 1999 Jul; 89(1): 181-6
6. Cameron AE, Sievert J, Asbury AJ, Jackson R, Gas leakage and the laryngeal mask airway. A comparison with the tracheal tube and the facemask during spontaneous ventilation using a circle breathing system, Anesthesia 1996 Dec;
51(12):1117-9
7. Lopez-Gil M, Brimacombe J, Cebrian J, Arranz J, Laryngeal mask airway in paediatric practice -A prospective study of skill acquisition by anesthesia residents, Anesthesiology 1996; 84:807-11
ADVANTAGES OF LOW FLOW ANAESTHESIA (LFA), FOREIGN GAS
ACCUMULATION, CONTRAINDICATIONS.
G. ROLLY, M.D., Ph.D.
Gent, Belgium The advantages
of low flow/closed circuit techniques are several fold: 1) diminished consumption of anaesthetic gases and cost containment 2) environmental pollution reduction (for workplace and atmosphere)
For foreign gas accumulation, successively will be dealt: 1) The sources of unwanted gas; 2) The individual compounds concerned and their clinical relevance; 3) How can they eventually be removed out of the system?
The foreign gases found in the anaesthetic circuit can artificially be divided in 4 components: 1) The compounds formed in the body: acetone, CO, methane and hydrogen; 2) The compounds absorbed in the body: ethanol, CO and nitrogen; 3) The compounds produced in the breathing circuit: CO and degradation products of the inhalational agents, particularly compound A; 4) The compounds introduced in the breathing circuit, as contaminants of medical gases, by returned anaesthetic gas and by inversed leaks: nitrogen and argon. Some identical compounds can be found in the different situations.
Normally nitrogen (N2) is present in the body. The normal content of the body, together with the lungs, amounts to about 2.7 l. During low flow/closed circuit anaesthesia sources of nitrogen can be: insufficient denitrogenation, use of side stream analysers, openig of a second soda lime canister still containing some nitrogen or a non airtight breathing system. If denitrogenation is performed with a high gasflow during 15-20 min, about 2 l of nitrogen is removed. The rest is slowly removed from less well perfused tissues and can give a nitrogen increase in the circuit during low flow conditions. The nitrogen can be 15 % and even higher during closed circuit. To eliminate nitrogen a high oxygen flow has to be given for at least 2-5 min. If no N2O is used there is no danger of hypoxia, but if a high concentration of N2O is used, lightening of anaesthesia has to be feared. Nitrogen can only be detected by mass spectrometry or a multigas analyser. In own research with rigid closed circuit anaesthesia mean concentrations of 10 % nitrogen were reached after 3 h, but individual values of 11 % were found.
Argon can accumulate in the circuit if in some exceptional circumstances oxygen concentrators are used. It is a noble and harmless gas. When long lasting low flow is used, 5-6 % argon has been described, and in minimal flow techniques 8-15 %. The argon can be washed out by a high gasflow every 90 min.
Hydrogen is washed out of the lungs at a rate of 0.6 ml/min. In a closed circuit it can accumulate by 200 ppm/h, well below flammable concentrations (4.6-94 % in O2; 5.8-86 % in N2O).
Carbon monoxide (CO) gives more concern as it is intensily bound to haemoglobin. Physiological values of COHb are around 0.4-0.8 %, but higher in smokers (even 10 %). In closed circuit Middleton reported that CO can increase with-in the range of 20-210 ppm; we found also a range of 7.5-164 ppm. Strauss found in closed system COHb values of 0.5-1.5 % in non-smokers and 3 % in smokers and Baum in minimal flow 1-1.5 % COHb. A toxicity index has been published by Henderson/Haggard: Itox = CO(ppm) x t(h); 600 ppm/h beginning intoxication, vomiting and headache at 900 ppm/h, life-threatening conditions at 1500 ppm/h. The normal endogenous production is 0.42