|
|
|
London 2003 |
|
| Page 7 of 7 |
INCREASING ANESTHETIC CONCENTRATIONS WITHOUT MODIFYING FGF BY SERIAL CONNECTION OF VAPORIZERS
Otero, P.E.1, DVM; Rebuelto,
M1, DVM, PhD; Hallu, R.E.1,
DVM; Aldrete J.A.2, MD, MS.
2Professor, Department of Anesthesiology,
University of South Florida, School of Medicine, Tampa, FL. President of Arachnoiditis Foundation, Inc.
When working with a vaporizer
located outside the breathing circuit the anesthetic vapour delivered into the
breathing system is dependent on fresh gas flow (FGF). For a constant FGF rate
the maximum amount of agent delivered into the breathing system depends on the
maximum output of the vaporizer employed. With a 300 ml/min basal FGF, a
maximum of 15 or 24 ml/min of anesthetic agent may be provided, if working with
a vaporizer maximal setting of 5 or 8%, respectively. As the requirements of an
adult patient (70 kg BW) during the first minutes of anesthesia are above the
mentioned, decreasing FGF from the beginning of the anesthesia will result on a
prolonged induction phase due to the low levels of delivered anesthetic. Other
disadvantage of using low flow anesthesia is the difficulty in producing rapid
changes in the anesthetic effect without FGF adjustments, due to the long time
constant of the low flow circuits. Flow-independent delivery of volatile agents is an
excellent alternative for this purpose. There are two possible approaches to
making the volatile agents delivery independent of FGF
i.e., placing the vaporizer inside the circuit (VIC) or injecting a calculated
volume of the liquid agent directly into the circuit. Lastly, even though the
use of vaporizers with a maximum output of volatile agent higher (3- or 4-fold)
than the currently used vaporizers seems reasonable, safety recommendations may
delay the use of such systems.
The purpose of this study was to
design a vaporization system capable of increasing the anesthetic
concentrations delivered to the breathing circuit without modifying the FGF
maintaining de vaporizer outside de circuit.
Agent-specific, variable-bypass,
temperature- and flow-compensated
sevoflurane (Sevo-Wick, Muraco
Medical Co., Ltd, , Tokyo, Japan, n = 4) and isoflurane
(Vapor 19.1 Vaporizer, North American Drager,
Telford, Pennsylvania, n = 4) vaporizers were used in this study. Vaporizers were calibrated to deliver 5%
maximum sevoflurane and isoflurane.
The same oxygen flowmeter
calibrated in 100 mL/min increments for flows from 0
to 1000 mL/min was used all over this experience.
Oxygen flow was started 30 minutes before measurements were done in order to
avoid temperature changes.
Anesthetic concentrations (vol%) were determined by placing the sampling head from an
anesthetic gas analyzer (Oxyanga Eku,
Leiningen, Germany) previously calibrated for
accuracy following manufacturer's
instructions into a corrugated tube (22 mm internal diameter, 60 cm length).
Corrugated tubing was used as reservoir when FGF was less than the necessary
volume for gas detection (150 mL/min). Each vaporizer
was tested at 100, 200, 300, 500 and 1000 mL/min at
its maximum tap setting for defining the actual output of anesthetic agent. The
study was divided in two phases. Sevoflurane- (S1, S2, S3 and S4) and isoflurane- (I1, I2, I3 and I4) vaporizers were studied on
phase 1 and 2, respectively. Vaporizers of each phase were connected to each
other, increasing in one each time. Thus, the first vaporizer (S1 or I1,
respectively) receives FGF from the oxygen flowmeter,
the second vaporizer (S2 or I2, respectively) receives as FGF the output of the
first vaporizer, the third vaporizer (S3 or I3) receives the output of the
second vaporizer, which results from the delivery of first and second vaporizer
(S1+S2 or I1+I2), and the fourth (S4 or I4) the addition of the first, second
and third vaporizers (S1+S2+S3 or I1+I2+I3) as delivered by the third
vaporizer. Thus, four series were described on each phase: S1, S1+S2, S1+S2+S3,
and S1+S2+S3+S4 for phase 1; I1, I1+I2, I1+I2+I3, and I1+I2+I3+I4 for phase 2.
Anesthetic agent concentrations
measurements for each series and each phase were measured at maximal vaporized
settings (5%) and oxygen flow rates of 100, 200, 300, 500 and 1000 mL/min. Each
measurement was repeated 6 times and between measurements the vaporizers were
turned off and the oxygen flow rate increased until no anesthetic agent
concentration was detected in the circuit. Differences between expected
(defined as the simple addition of the maximum values previously determined for
each vaporizer) and measured values were calculated for each series of all
phases. The ambient temperature in the working box was constant throughout the
entire study. Box waste gas was eliminated by a gas-scavenging system.
Results are reported as mean ± SD. An analysis of variance was used to
determine whether differences existed among means of anesthetic concentrations
at each series of each phase at different FGF rates. When differences were
found, a Tukey’s Multiple Comparison test was done
for determining which means were different. Significance was set at a 5% level.
Results
Anesthetic agent concentration measurements for each series of each phase are shown on Table 1 and 2. Statistically significant differences were detected for series of both phases for 100 mL/min FGF and in phase 2 for 200 mL/min (P < 0.001).
The differences between expected and
measured concentrations are listed on table 3.
Table 1. Anesthetic agent
concentration measurements for each series of phase 1.
|
Phase 1 (Sevoflurane) |
||||
|
FGF (ml/min) |
S1 (n = 24) (vol%) |
S1+S2 (n = 6) (vol%) |
S1+S2+S3 (n = 6) (vol%) |
S1+S2+S3+S4 (n = 6) (vol%) |
|
100 |
4.3 ± 0.40* |
7.3 ± 0.09* |
10.8 ± 0.05* |
13.2 ± 0.08* |
|
200 |
4.6 ± 0.38 |
8.2 ± 0.08 |
11.4 ± 0.10 |
13.9 ± 0.29 |
|
300 |
4.7 ± 0.29 |
8.3 ± 0.05 |
11.6 ± 0.16 |
14.1 ± 0.23 |
|
500 |
4.8 ± 0.31 |
8.3 ± 0.05 |
11.6 ± 0.18 |
14.1 ± 0.08 |
|
1000 |
4.7 ± 0.26 |
8.3 ± 0.11 |
11.3 ± 0.24 |
14.1 ± 0.16 |
* statistically
significant differences (P<0,001) between FGF rates
Table 2. Anesthetic agent
concentration measurements for each series of phase 2.
|
Phase 2 (Isoflurane) |
||||
|
FGF (ml/min) |
I1(n =24) (vol%) |
I1+I2 (n = 6) (vol%) |
I1+I2+I3 (n= 6) (vol%) |
I1+I2+I3+I4(n= 6) (vol%) |
|
100 |
4.2 ± 1.32* |
7.4 ± 0.20* |
10.3 ± 0.10* |
13.1± 0.08* |
|
200 |
4.2 ± 0.91* |
7.5 ± 0.20* |
10.4 ± 0.15* |
13.0 ± 0.07* |
|
300 |
4.8 ± 0.52 |
8.3 ± 0.05 |
12.4 ± 0.08 |
15.5 ± 0.08 |
|
500 |
4.9 ± 0.32 |
8.5 ± 0.05 |
12.5 ± 0.18 |
16.1 ± 0.08 |
|
1000 |
4.9 ± 0.27 |
8.5 ± 0.06 |
12.5 ± 0.16 |
16.1 ± 0.08 |
* statistically
significant differences (P<0,001) between FGF rates
Table 3. Differences
between expected and measured concentrations for both series.
|
Values |
S1+2 |
S1+2+3 |
S1+2+3+4 |
I1+2 |
I1+2+3 |
I1+2+3+4 |
|
Expected |
9.2 |
13.8 |
18.4 |
9.2 |
13.8 |
18.4 |
|
Measured |
8.1 |
11.3 |
13.9 |
8.4 |
11.6 |
14.8 |
|
Expected-Measured |
1.1 |
2.5 |
4.5 |
0.8 |
2.2 |
3.6 |
Discussion
Our results show that the delivered final anesthetic agent concentration after connecting serially 2, 3 and 4 vaporizers increased with each vaporizer addition, however, the final output was lower than the expected one. This difference may be due to the fact that carrier gas composition for vaporizers 2, 3 and 4 contains partly the anesthetic agent. Thus, anesthetic vapour of the carrier gas flow mixes with anesthetic present in the vaporization chamber, were saturation pressure exists. This results in the impairment of the carrier gas capability of vaporization. The decrease on final concentration is repeated along vaporizers and the influence is higher as the number of connected vaporizers increases. This behavior was similar for both sevoflurane and isoflurane, as they share the same vaporization method and they have similar physical properties. The difference found for 100 mL/min FGF for sevoflurane and 100 and 200 mL/min FGF for isoflurane were expected, as those flows are outside the range of calibration of the used vaporizers.
Approximation to final anesthetic concentrations may be calculated by
means of equation 1:
![]()
[1] % vap 1+2+3+4 =
(Fl vap S 1-4) – (FL vap
Ex S 2-4) x 100
FGF + (Fl vap S 1-4) – (FL vap
Ex S 2-4)
where:
·
FL vapS 1-4 = final amount of the vapour delivered by each
vaporizer.
·
FL vap Ex S 2-4, = final amount of the vapour extracted by each
vaporizer
Equation 1 is resolved as follows:
![]()
[2] FL vap1 = FGF x
%V1
100- %V1
![]()
![]()
[3] Vcarrier2 =
((FGF + FL vap1) x %V2)
x 100
(100- %V2) %CVC
![]()
[4] FL vap Ex2 = Vcarrier2 x %V1
100
![]()
[5] FL vap2 = (Vcarrier2
– Fl vapEx2) x %CVC
100
![]()
[6] % vap1+2 =
(Fl vap1 + Fl vap2 – Fl vapEx2) x 100
FGF + (Fl vap1 + Fl vap2
– Fl vapEx2)
![]()
![]()
[7] Vcarrier3 =
((FGF + FL vap1 + Fl vap2 – Fl vapEx2)
x %V3) x 100
(100- %V3) %CVC
![]()
[8] FL vap Ex3 = Vcarrier3 x %vap1+2
100
![]()
[9] FL vap3 = (Vcarrier3
– Fl vapEx3) x %CVC
100
[10] % vap1+2+3 =
(Fl vap1 + Fl vap2 +Fl vap3 – Fl vapEx2
- Fl vapEx2) x 100
FGF + (Fl vap1 + Fl vap2 +Fl vap3– Fl vapEx2 - Fl vapEx2)
![]()
![]()
[11] Vcarrier4 =
((FGF + FL vap1 + Fl vap2+ Fl vap3 – Fl
vapEx2 - Fl vapEx2) x %V4) x
100
(100-%V4) %CVC
![]()
[12] FL vap Ex4 = Vcarrier4 x %vap1+2+3
100
[13] FL vap4 = (Vcarrier4
– Fl vapEx4) x %CVC
100
[14] FL vapS 1-4 = [2] + [5] + [9] + [13]
[15] FL vap Ex S 2-4 = [4] +
[8] + [12].
Where:
FGF (mL/min)= initial fresh gas flow
Vcarrier n (mL/min)= Volume of carrier gas flow and n = series
location
%Vn
(%)= tap setting of vaporizer and n = series location
%CVC (%)= vapour concentration in the vaporization
chamber
FL vapn
(mL)= vapour flow added by
each vaporizer and n = series location
FL vapExn
(mL)= vapour flow extracted
on each vaporizer and n = series location
Figure 1 shows expected, measured and calculated (as equation 1) values.
Figure 1. Expected, measured and calculated
(as equation 1) values.
Conclusions
In this study we demonstrated that
anesthetic agent concentrations may be increased by serially connecting
agent-specific vaporizers without modifying FGF and keeping the vaporizers
outside the breathing system, and we propose an equation for predicting the
approximate delivered anesthetic concentrations. This technique may be an
alternative in low flow anesthesia, as provides constant and predictable final
anesthetic concentrations.
References:
1.
Aldrete JA,
Lowe HJ, eds. Low flow and closed system anesthesia. Grune
& Stratton, New York, 1979.
2.
Baum Jan. Low Flow Anesthesia: the
theory and practice of low flow, minimal flow and closed system anesthesia. 2nd
ed. 2001. Butterworth-Heinemann
3.
Lowe
HJ, Ernst EA. The quantitative practice of anaesthesia. Use 0f closed circuit. Willians and Wilkins eds. Baltimore 1981
4.
Nunn
G. Flow independent delivery of volatile agents. Lecture. ALFA 2002, Pisa,
26-27 April 2002
How
long should be high flow phase in children for equilibration of sevoflurane
prior to low flow anaesthesia when oxygen used as carrier gas?
P .
Bozkurt, M Bakan, E Tomatir*, Ö Sen, N Kurt, M Posta, G Kaya
İstanbul University, Cerrahpasa
Medical Faculty and Pamukkale University Medical
Faculty*, Departments of Anaesthesiology, Turkey
Avoiding
use of nitrous oxide had several advantages as being ecological and ease of
teaching low flow anaesthesia to juniors (1). The literature about performance
of low or minimal flow anaesthesia in children is limited (2). The anatomical
and physiological differences of children bring up special management
techniques also for low flow anaesthesia.
The
aim of this study was to find out whether the high flow initial phase could be
shortened in children when oxygen used as carrier gas with sevoflurane prior to
low flow anaesthesia.
Material
and Method: Following institutional ethics committee approval 49 children who were
free of respiratory diseases, anemia, low cardiac
output or difficult intubation risk, ages ranging
from 1day to 8 years included in this study. Children who had an intravenous
line in were given thiopentone 5mg/kg or propofol 2mg/kg and atracurium or
cis-atracurium for induction and remifentanil
infusion at a rate of 0.5 mg/kg/min started. Mask
ventilation with a tight fitting face mask was started immediately using a semiclosed circle absorber system. Sevoflurane 2-4vol % and 100% O2
initiated at a flow of minute ventilation of the child and ventilated manually
until the neuromuscular blocker drug effect achieved. Remifentanil
infusion rate lowered to 0.25mg/kg/min 4 min later The
children were intubated with endotracheal tube which is appropriate for their
age by a single attempt. The tube
secured immediately without disconnecting the tube and the Y piece and
mechanical ventilation started with the settings tidal volume 8mL/kg and
respiratory rate according to age. A
single anaesthesia machine (Cicero EM Drager-Germany)
used during the study. The flow rate of O2, respiratory rate, presence of air leakage
were recorded. Starting from the placement of the mask to the patients
face the FiSevo ( inspired
fraction of sevoflurane), FiO2 ( inspired fraction of oxygen) FetSevo (endtidal fraction of
sevoflurane)and ), FetO2 ( endtidal
fraction of O2) were recorded every minute and until intubation
and for 10 minutes long after the intubation from the
monitor of Cicero EM anaesthesia machine. The initial sevoflurane ratio and
flow of O2 were kept constant throughout the study period. The time to equilibration between inspired
and end tidal concentrations of sevoflurane and the concentrations at
equilibrium were recorded. Equilibration is defined as reaching the FetSevo level reaching to 80% of FiSevo
level. Later the composition of the gases changed to 50% oxygen in air for the
rest of the anaesthesia. The patient
records were grouped according to their ages as children less than 2 years old
and 2-8 years old groups.
The
results were given as mean ± standard deviation. Between
groups comparisons were performed with Student’s t test. P value less than 0.05
considered significant.
Results:
Twenty-one children were less than 2 years old and 28 were older than 2 years
of age. The demographics and ventilatory settings of the groups are given in
Table 1.
|
|
Age ( years) |
Weight (
kg) |
Resp. Rate (breaths/min) |
O2
flow( L) |
Duration
of mask ventilation (min) |
|
<2
years (min-max) |
1±0.8 1day-2years |
9.2±4.2 2.5-15 |
22.4±5.9 17-30 |
1.6±0.5 0.8-2.5 |
2.8±0.7 2-4 |
|
2-8
years (min-max) |
5.5±2.5 2-8 |
19.4±7.7 7-24 |
17.6±1.7 20-14 |
2.8±0.9 1.4-3.7 |
2.9±0.8 1.5-4 |
None
of the patients were hypoxic during the manual ventilation by mask ( SpO2 >94%). SpO2 ranged between
99% and 100% following intubation until the end of
the study. In children less than 2 years old the equilibration between inspired
and end tidal sevoflurane concentrations reached at 2.8±
2.5 min after induction. In the older children this period was 7.1±2.5
min. The difference between age groups was statistically significant ( p=0.0001). At this point inspired sevoflurane
concentration was 2.3± 0.8vol % for both groups.
The end tidal sevoflurane concentration was 2±0 .8vol % and 1.9±
0.6vol % for patients less than 2 years old and the older group respectively.
Discussion
and Conclusion: The applicability of low flow and minimal flow techniques are increasing
with the advent of modern anaesthesia machines also in children. Because the children are more prone to hypoxia during induction
pure oxygen used. The tendency of using oxygen as carrier gas encouraged
us for setting up this study. Before
decreasing the flow the uptake of inhalation agent required to reach to an
equilibration for avoiding superficial anaesthesia. In children less than 2
years old duration is fairly short when compared to older children. This was
probably dueto high alveolar ventilation and
distribution of cardiac output and less blood-gas solubility (
by the influence of hematocrit, hemoglobin type and plasma protein fraction) (3). Baum had
stated an interval of 10 minutes for high flow required for adults when N2O
use omitted (1). The duration of high
flow (in our setting high flow = minute ventilation) before the start of low
flow could be decreased in small children. Low flow techniques could be started
in a few minutes without risk of development of hypoxia or awakening when
oxygen used as carrier gas for sevoflurane. This method also provides ease of
teaching of low flow anaesthesia and could be applied safely in shorter cases
while switching to low flow is more rapid.
References:
Presenting
author:
Pervin Bozkurt, M.D., Associate
Professor of Anaesthesiology
İstanbul University Cerrahpaşa
Medical Faculty
Department
of Anaesthesiology
34303 İstanbul- Turkey
Tel:
90 212 560 81 87
Fax:
90 212 529 56 00
E-mail:
apbs@istanbul.edu.tr or apbs@turk.net
NAGARAJAN.R+ MD,RAMPRABU.K* MD,BENJAMIN NINAN+
+ Intitute of cardiovascular diseases, Madras medical mission ,Chennai
*PGIMER Chandigarh,
BACKGROUND
In the present day practice of cardiac anaesthesia inhalational agents and propofol have replaced opioid based anaesthesia during cardiopulmonary bypass(CPB).Inspite of this shift in practice the same open method of administration of inhalational agents is being used which results in significant wastage of these agents.The need for a closed circuit CPB ventilator is further necessitated by the now common ,normothermic CPB which requires higher blood flow rates and therefore a correspondingly higher fresh gas flows.For example, with the present day system for an average Indian of 1.6m2 ,at normothermia,the blood flow requirement would be 3.8 l/min(2.4 l/m2 )and corresponding fresh gas flow of 2.3 to 3.8 l/min(0.6 to 1 ratio of fresh gas flow and blood flow).
Herein we present a patented design of closed circuit CPB ventilator which would enable use of metabolic fresh gas flow during CPB.
The three major problems which have to be addressed in designing a closed circuit CPB ventilator are
1.Oxygenators lack the elasticity of lungs
2.The flow needs to be continuous through the oxygenator, any interruption of fresh gas flow would constitute a shunt flow.This is because of the lack of adequate functional residual capacity of the oxygenators.
3.The pressure on the gas side of the blood gas interface should be less than 10cmH2O ,any increase in pressure above this would increase the possibilty of air embolism.
All these problems have been cicumvented in this design by using two ventilators in series.The system is based on the principle of one ventilator ventilating the other in a sequential manner.The functional capabilities of these ventilators would have to be different from the available conventional anaesthesia ventilators.The present day tecnology would allow for such a modification to be made with ease.As a safety feature this design also incorporates a pressure limiting valve and a pressure protection valve
Use of metabolic fresh gas flows even during CPB would be possible using such a system.
Opening
the Horizons for Respiratory Xenon Anaesthesia:
This paper has been removed at the author's request.