|
|
|
San Sebastian 2004 Session 1-3 |
|
|
<Previous page> <Back to Abstract Home Page> <Next Page> <Program> |
Summary
"Low-flow
anaesthesia" is a term applied to techniques in which fresh gas flows are
less than the alveolar ventilation used. Depending on the specific technique,
fresh gas flows may vary between 30 mL/min and 200 mL/min for an adult. Methods
of anaesthesia with reduced fresh gas
flow, including the technique of “quantitative anaesthesia” with a completely
closed rebreathing system have gained more and more interest in recent years.
The potential for continuous and comprehensive analysis of anaesthetic gas
composition, mandatory safety pharmacokinetics and pharmacodynamics of the
inhalation anaesthetics justify the revival of these methods in veterinary
anaesthesia.
Introduction
In the last years, there have
been significant advances in veterinary anesthesiology. However, due to
economic reasons the incorporation of new drugs has been deleted. Although this
reality affects most countries around the world, in Latin America, the
situation becomes more critical with time and thus becoming more necessary to
resort to new techniques leading to greater efficiency. In the context of
inhaled anaesthesia, the use of circular circuits appears to be a practical and
safe alternative. The possibility of making the patient rebreathe part of or,
even better, of all the totality of the exhaled gases decreases the anaesthetic
vapour intake and therefore reduce the consumption of the selected agent. When
working with low flow, even without completely closing the circuit, heat (1)
and humidity (2) are conserved within the circuit, improving the clinical
condition of the anaesthetized patient. A reduction in the anaesthetic gas flow
would also leads to a lower level of contamination within the operating room
and onto the environment. Although the amount of anaesthetic gases delivered to
the atmosphere as a result of veterinary anaesthesia is insignificant compared
to the amount release from human operating rooms, it is crucial to take into
consideration the growing pollution problem and thus consider the improvement
of environmental conditions. The use of anaesthetic circuits that allow the
rebreathing and the reduction of fresh gas flows are as old as inhaled
anaesthesia itself (3), but its incorporation to the practice is nowadays a
reality in anaesthesiology. It allows the incorporation of new generation
agents to routine procedures without significantly modifying the cost of the
anaesthesia, it increases the advantages of hemodynamic stability, it
demonstrates a marked analgesic effect and reduces the induction and recovery
times. Although it is difficult to incorporate low flow techniques into the
routine- veterinary practice due to the lack of infrastructure, we will try to
define the basis for a safe and predictable procedure.
Oxygen and low flow
The total oxygen intake as
well as the diluted ratio of the gas
mixture inhaled by the patient have to be taken into account when designing low
flow anaesthesia. The circle rebreathing circuits can operate half-opened,
half-closed or completely closed, varying in the level of exhaled gases that
are re-inhaled and the amount of fresh gas administered. It is generally
accepted that in low flow anaesthesia the level of rebreathed gases are above
50% (4). This varies with the size of the animal and the circuit volume. The
primary obstacle in the use of low flow techniques is that the traditional
vaporizer is flow dependent. Thus, working with small animals, the flow needed
to maintain an appropriate level of re-inhalation is out of the denoted range. On the other hand while working with large
animals, it is not possible to reach an adequate amount of anesthetic vapor
when using a fresh gas flow (FGF) similar to the animals metabolic O2
consumption due to a preset maximum
vaporizing level of 5 %. This observation
occurs during the first 30-40 minutes of anesthesia which is where the greatest
uptake of anesthesia occurs.
It is extremely important to
ensure an oxygen supply that fulfils metabolic needs in order to reach an
adequate inhaled rate (FiO2) and thus avoiding hypoxemia.
The metabolic oxygen consumption is highly influenced by the patient’s weight,
its body size, its body temperature, the degree of the CNS depression and the
anesthetic agent. A quick and simple method to calculate the oxygen consumption
is the formula suggested by Brody (5) for all homoiotherms:
VO2
= 10,15 x PC(kg)0,75 (mL/min) (Equation 1)
Where, VO2 is the
oxygen uptake (mL/min) and BW is the body weight (kg).
According to the studies
performed by Lowe (6), in patients depress by anaesthetics, the oxygen
consumption is about 30 % lower than that calculated by Brody’s formula.
Studies performed in anaesthetized animals suggest that an oxygen flow of 3 to
4 mL/kg/min is enough to reach the metabolic oxygen requirements during the
anaesthesia in small (7) as well as in large animals (8).
When the reservoir bag is
inside the circuit, an easy and simple way of adjusting the FGF is by providing
a fresh gas volume to maintain the bag’s volume unchanged. An increase in the volume
of the bag would involve a gas supply greater than what the animal intakes, yet
a decrease in the volume is associated to an insufficient fresh gas supply. It
is assumed that the volume of gases added to the circuit equals the volume of
gases incorporated by the patient plus the concomitant leaks.
Depending on the equipment,
the evaluation could be complicated by the use of artificial ventilators. When
working with a descending (during expiration) bellow system where the
accumulation of excess gas can not occur, the airway pressure must be under
control at all time. Sub zero pressures at the moment of inhalation is a result
of a contraction of the system due to insufficient FGF, while positive
pressures denotes an excess in FGF. The
latter, only happening when the ventilator pop off valve is closed. When
working with an ascending (during expiration) bellow system, the bellow is
maintained between 100 and 200 mL above the base. Therefore, changes in the
height of the bellows will show a deficit or an excess of fresh gas, which can
be easily detected and corrected by adjusting the FGF. To guarantee enough
oxygen supply and a correct oxygenation of arterial blood, a FiO2
of at least 30% must be maintained during the totality of the procedure. When
oxygen is the carrier gas, it is unlikely that a hypoxic mixture is generated,
even if extremely low flows are used and more so if the circuit has a reservoir
bag. In large animals, the recumbency is frequently associated to severe
ventilation disorders. In these patients, hypoxemia is a frequent complication
that demands ventilation control and a FiO2 greater than
the percentage previously suggested. The use of oxygen as a unique carrier gas
is considered more and more as a better option (9). The FiO2 in
circuits that work with low flows begin with O2 levels close to 100%
but after approximately 5 to 10 minutes this percentage decreases to about 60%
due to the elimination of N2 accumulated in the body.
Numerous studies coincide in
concluding that a hyperoxemic mixture yields a high security margin. The
incidence of atelectasia observed in patients exposed to high and low oxygen
concentrations is not statistically significant (10), while the activity of the
immune system is improved (11), as shown by a lower incidence of post-surgery
infections in patients who received O2 as a unique carrier gas.
Respiratory complications are also less frequent in patients inhaling high
concentrations of O2 during the anaesthetic procedure (12). In
patients with no lung damage, the potential adverse effects that may result
from a high exposure of O2 would only be present after many hours,
but this is extremely uncommon in everyday practice (13).
The use of nitrous oxide is
not commonly utilized in veterinary medicine. This gas has a low potency in
most animal species (MAC above 200%). In addition, the use of low solubility
anaesthetics such as isoflurane, sevoflurane or desflurane, avoids the
collateral effect of nitrous oxide as a “second gas”. The presence of nitrous
oxide in the anaesthetic mixture of a circuit working with low flows demands
rigorous monitoring due to variations in its uptake along the anaesthetic
procedure.
When planning low flow
anaesthesia, it is important to know the pharmacokinetics of anaesthetic gases
and the rate between their pressures in different sections of the system. The
objective of a general anaesthetic is basically to reach an adequate
concentration in the CNS in order to perform different surgical maneuvers
without producing pain or movements. The volume of the anaesthetic vapour to
achieve this purpose is introduced into the body through the respiratory
system. In the lungs, the blood collects the anaesthetic and carries it to
different tissues, including the CNS. The magnitude and speed of this process
depend on some factors. The most important parameter that conditions the
kinetics of anaesthetic gases is the volume of alveolar ventilation and the
alveolar concentration of the anaesthetic. The concentration gradient through
the alveolar-capillary membrane and the solubility of the agent will determine
the partial pressure of the anaesthetic in arterial blood. Therefore, it is
necessary to take into account, during the totality of the procedure, the
relationship between the variables that determine the concentration of the
agent in the body.
Lowe, based on Züntz and
Severinghaus studies, developed a mathematics equation to calculate the
anaesthetic uptake in each anaesthetic phase (14).
VAN = Ca x Q x t –½ (Equation 2)
Where VAN is the total anaesthetic uptake, Ca is its arterial concentration, Q is
the cardiac output and t is the time.
The arterial concentration can
be calculated multiplying the alveolar concentration (CA) by the
blood/gas solubility coefficient of the agent (lB/G)
Ca = CA x lB/G
(Equation 3)
Finally, the alveolar
concentration can be calculated via the following equation,
CA
= f x MAC
(Equation 4)
Where MAC is the Minimum
Alveolar Concentration of the selected anaesthetic in that species and f is the
fraction of the MAC (Table 1).
DRUG |
Canine |
Feline |
Equine |
Bovine |
Ovine |
Swine |
|
Halothane |
0.87 |
1.19 |
0.88 |
0.76 |
0.97 |
0.91 |
|
Isoflurane |
1.28 |
1.61 |
1.31 |
____ |
1.58 |
1.45 |
|
Enflurane |
2.06-2.2 |
2.4 |
2.12 |
____ |
2.0 |
____ |
|
Sevoflurane |
2.34 |
2.58 |
2.34 |
____ |
3.3 |
1.97 |
|
Desflurane |
7.20 |
9.80 |
7.23 |
____ |
9.5 |
10.0 |
|
Nitrous oxide |
188-200 |
150 |
190 |
223 |
____ |
195-277 |
|
Metoxiflurane |
0.29 |
0.23 |
0.22 |
0.26 |
0.26 |
____ |
Table 1: MAC of volatile anaesthetics (%vol).
* One time the MAC produces a mild anaesthesia, 1.5 times the MAC
produces a moderate surgical anaesthesia, two times the MAC, produces a deep
anaesthesia.
As can be inferred from the
equation, when there is a constant anaesthetic supply, the uptake vs. time
curve describes a decreasing exponential curve (Figure 1)

Figure 1: Figure A; isoflurane uptake (FA = 1.3 %vol) in a
510 kg B.W. horse, calculated by means of Lowe’s formula. Figure B; cumulative
isoflurane doses in the same patient.
As can be observed in figure
1, The uptake of anaesthetic during the first minute of anaesthesia is equal to
the uptake in different periods (minutes 1 to 4, minutes 4 to 9 and minutes 9
to 16) and it continues indefinitely until it reaches a theoretical balance,
which is difficult to reach in a typical anaesthesia.
The dose to be administered
after the first minute was named by Lowe as “unit dose” (UD), which is
expressed in mL of anaesthetic vapour and can be calculated by means of the
following equation,
UD
= 2 x Ca x Q (Equation 5)
The unit dose represents
double the volume of vapour circulating in the system and must be administered
at 1, 4, 9, 16, 25, 36 etc. minutes, to maintain an anaesthetic supply that
keeps the alveolar concentration at a stable level.
The amount of anaesthetic
vapour uptaken by the tissues can be calculated by multiplying the UD by the
square root of time. This is known as cumulative dose (CD).
CD = DU x Öt
+ c (Equation
6)
where c represents the
arterial concentration after the first dose.
At the beginning of the
anaesthesia, the dose should be calculated considering the volume in which the
anaesthetic will have to be diluted in order to reach equilibrium quicker. This
dose is called “initial” or “priming dose” (PD) and is calculated by the
following equation,
PD
= CA x (VS + VL) + Ca x Q
(Equation 7)
Where VS represents
the volume of the anaesthetic circuit and VL is the volume of lungs
and airways. In both cases, these volumes are expressed in deciliters (dl).
It
is essential to use these equations along with the volumetric addition of the
anaesthetics in closed circuits. It is important to emphasize that there are
some objections to Lowe’s theory. Lin suggests to keep closed circuit anaesthesia
by means of the volumetric addition of some anaesthetics. Lung uptake of the
anaesthetic agent is constant, at least during the first 2 hours of
administration. Therefore, in a 70 kilogram individual, after an initial period
of high uptake related to the central compartment saturation, the halothane
uptake is maintained between 15 and 20 ml/min of vapour for each 1% of
anaesthetic in the FA. For isoflurane and enflurane, this rate is
reached with 10-15 ml/min and 30 ml/min, respectively. Eger also suggests that
the desflurane uptake is constant during anaesthesia (34). Studies performed in
animals by Prof. Moens et al. (35),
demonstrated that alveolar sevoflurane concentration shows an evident
decreasing tendency when Lowe’s administration diagram is used, suggesting that
uptake does not decrease as a function of the square root of time, as occurs
with other anaesthetics like halothane e isoflurane (8).
The low flow technique always
obliges one to constantly have in mind the uptake diagram. One should try to
reach equilibrium between the volume of the anaesthetic vapour supplied and the
uptake of this volume, creating a procedure extremely efficient. Therefore, if
the O2 supply is adjusted constantly during the anaesthesia
procedure, a quantitative anaesthesia can be obtained.
Fractions
It is extremely important to
know the ratio between the selected agent and the anaesthetic mixture in each
part of the system. The fractions to take into account during anaesthesia are
the vaporized fraction (FV), the inhaled fraction (Fi)
and the alveolar fraction (FA).
The FV represents the volume of vapour released by
the anaesthetic machine and it is the result of the dilution between the column
of gas that goes through the vaporizer without contacting the anaesthetic
(diluent flow) and the column that goes into the vaporizing chamber to collect
the anaesthetic (vaporizer flow). Most of the vaporizers have a maximum
vaporizing limit that, for security reasons, is about 4-5 times the MAC of the
anaesthetic agent. Thus when working with an agent specific vaporizer, the
volume of vapour added to the system is the result of multiplying the FGF,
expressed in deciliters (dl), by the percentage delivered by the vaporizer.
Thus, if we work with an FGF of 0.5 liters/min with the dial at 5%, we will be
releasing 25 mL of anaesthetic vapour per minute. This is valid for any
anaesthetic agent.
When working with VOC, the
volume of anaesthetic vapour to be released is a function of FGF and it is
limited by the vaporizing maximum of the equipment. It is quite difficult to
reduce the FGF during long uptake periods or in large animals.
The Fi represents
the anaesthetic concentration inhaled by the patient, expressed in volume
percentage (%vol). When working with non-rebreathing circuits, the Fi
is equal to the FV. However, when working with circuits that allow
rebreathing, the Fi is the result of the mixture between the
rebreathing gas column and the fresh gas column. Therefore, the final
composition of the mixture will depend on how much anaesthetic vapour is
delivered into the system and how much remains within the system after each
uptake period. It is important to keep in mind that, as the FGF is reduced, the
number of the vaporizer dial must be increased to maintain a constant anaesthetic
supply.
The FA or
anaesthetic alveolar concentration represents the percentage of vapour that
returns to the lungs with venous blood and it is closely related to the partial
pressure of the anaesthetic in tissues, including the CNS. The FA is
always smaller than the Fi due to the tissue uptake process. The
ratio (FA/Fi) basically depends on the blood/gas
solubility coefficient of the agent and it is smaller when the solubility is
smaller. The magnitude of alveolar ventilation and the cardiac output also has
influence on this gradient (15).
The ratio FA/Fi
is independent on the anaesthetic system because once the drug enters the body,
its kinetics is only influenced by factors that does not depend on the
anaesthetic technique (14).
Semi closed Circuits
The nomenclature clearly
defines the different anaesthetic systems based on total FGF in human medicine
(Table 2).
|
|
Oxygen(L/min) |
N2O(L/min) |
Total (L/min) |
Author |
|
Intermediate
flow |
1.0 |
1.0 |
2.0 |
Aldrete
& Romo (16) |
|
Low
Flow |
0.5 |
0.5 |
1.0 |
Foldes
& col. (17) |
|
Minimum
flow |
0.3 |
0.2 |
0.5 |
Virtue
(18) |
|
Almost closed circuit |
0.5 |
---- |
0.5 |
Aldrete
(19) |
Table 2: suggested nomenclature to
classify systems with more than 50% of rebreathing.
On the other hand, to classify
the system in veterinary medicine, the FGF must be related to the patient’s
weight (table 3).
|
|
Oxygen (mL/kg/min) |
N2O (mL/kg/min) |
Total (mL/kg/min) |
Author |
|
Semi
closed system |
21 |
21 |
42 |
Muir
& Hubbell (20) |
|
Low
flow system |
10 a
15 |
--- |
10 a
15 |
Warner
& Bednarski (21) |
|
Closed
system |
4.4 a
6.6 |
--- |
4.4 a
6.6 |
Muir
& Hubbell (22) |
Table 3: flows suggested in veterinary medicine for different anaesthetic systems.
In summary, the percentage of
re-inhaled and how much anaesthetic vapour is being added to the circuit every
minute must always be kept in mind.
In patients who are well
ventilated and hemodynamically compensated, the desired alveolar concentration
will depend on the Fi, and the latter will depend on the mixture
between the fresh gas column and the rebreathing gas column.
The decrease in the
halogenated agent concentration in the circuit has two origins. One being the
tissue uptake and the other being the lost anesthetic gases, which are expelled
outside the system through the pop off or relief valve and leaks of the
circuit. No matter what the cause of the vapor loss, the same amount must be
replenished every minute to keep the Fi constant. If we multiply the
patient’s respiratory volume by the percentage of anaesthetic in the alveolar
and the inspired fractions, we will obtain the volume of vapour to be replenished
every minute. Therefore, if the patient’s respiratory volume is 2 liters/min
and has a FA of 1.5% and a Fi of 2%, the inspiration of
40 mL of anaesthetic vapour per minute and an expiration of 30 mL occurs (23).
Therefore, 10 mL of vapor should be added to keep the concentration constant.
In the same patient, if we were working with a vaporizer outside the circuit, a
FGF of 0.5 L/min. with the vaporizer dial at 2% would be enough to guarantee
the stability of a system with rebreathing of 100%. The same would happen with
a FGF of 0.25 L/min and the dial at 4% or a FGF of 0.1 L/min and the dial of a
hypothetical vaporizer at 10%.
Circuits that can work with
low flows, although not completely closed, create an easier manipulation,
primarily in patients that do not exceed 40-50 kg of body weight. The
possibility of using low solubility compounds like sevoflurane or desflurane,
whose vaporizers allow releasing a high percentage of anaesthetic vapour,
simplifies our job.
The use of an almost closed
circuit system is the most appropriate for a typical anaesthetic procedure
because it can be used without integrating new equipment. In order to avoid the
excess gas to increase the pressure of the circuit, one should leave the
expiratory valve open slightly. In these cases, the anaesthesia could be
planned as follows:
In order to calculate the
volume of vapour necessary to saturate the system at the desired concentration,
the first thing to know is the total volume of the system.
VT
= VS + FRC + VCC + VCB (Equation 8)
Where:
VT: is the total
volume of the system.
VS: is the volume
of the anaesthetic circuit.
FRC: is the functional
residual capacity (see table 4).
VCC: is the volume
of the central compartment (calculated as 10% of the patient’s body weight).
VCB: is the volume
of circulating blood (calculated as 60% of total blood volume of that species).
Aldrich and Haskins (24) estimated this volume at 9-10% of body weight in dogs
and 5% in cats.
Once the total volume of the
system and the desired anaesthetic concentration is known, the volume of vapour
to reach the desired concentration can be calculated. For instance, if the
patient weighs 40 kg and the circuit volume is 5 liters, the total volume will
be 13.4 liters. Therefore, if one desires 1.5% of anaesthetic vapour diluted in
13.4 liters, we must add 201 mL of anaesthetic vapour. If the vaporizer
releases a maximum of 25 mL of vapour with a FGF of 5 mL/min, the system will
be loaded with the desired concentration in 8 minutes. Once the system is
saturated and the desired anaesthetic depth is achieved, the induction or
impregnation phase is completed and the maintenance phase begins.
The vaporization level has to
be simultaneously adjusted by regulating the FGF and the vaporizer dial, taking
into account the tissue uptake and the anesthetic leaks. To estimate this, it
is necessary to calculate the vapor deficit every minute. Although these
calculations could only be made by means of halogenated gases analyzers, the
relationship between the anesthetic level and the parameters used when
monitoring the patient allows the utilization of this diagram with confidence.
Regardless of the system used,
keeping the patient in a light anesthetic plane by performing a balanced
anesthesia, decreases the amount of vapor to be added and thus increases the
safety of the procedure.
The minimum flow level of the vaporizer has to be taken into
account to avoid mistakes. In large animals, when working with vaporizers
outside the circuit with a low vaporization limit (less than 20%), the
reduction in FGF to metabolic levels will lead to drug deficit. This will
prevent us from keeping the patient in an adequate anesthetic plane. The
situation becomes more critical when the anesthetic agent has a high solubility
and/or the equipment has a low vaporization limit. In these cases it is
advisable to use methods that supplies the anesthetic vapor independently from
the FGF (25). The volumetric addition has proved to be an effective and safe
method to deal with this difficulty.
|
Specie |
Canine |
Feline |
Equine |
Bovine |
Goat
|
Sheep
|
Swine |
Rat
|
|
Weight (Kg) |
---- |
3-4 |
400-500 |
400-500 |
35-45 |
30-40 |
12-15 |
0,1-0,2 |
|
BF (rpm) |
13-25 |
20-25 |
8-10 |
20-30 |
12-15 |
35-40 |
10-15 |
80-100 |
|
TV (mL/kg) |
15-20 |
8 |
10,1 |
7,1 |
12,9 |
8,3 |
10-15 |
7-8 |
|
VMR(mL/min) |
---- |
550-700 |
74600 |
85977 |
6300 |
10400 |
2700 |
80-200 |
|
VMR(mL/kg/min) |
200 |
174 |
154 |
166 |
174 |
297 |
200 |
650-700 |
|
CFR (mL/kg) |
53,6 |
17,8 |
36,6 |
31,9 |
49,6 |
45,3 |
---- |
6,8 |
|
RV (mL/kg) |
16,7 |
---- |
19 |
16,1 |
---- |
---- |
---- |
4,2 |
TABLE 4: Respiratory parameters in different species (approximate values).RF: Breathing frequency TV: Tidal Volume; VMR: Volume Minute Respiratory ; CFR: Functional Residual Capacity; RV: Residual Volume.
Closed
Circuits
According to Lowe, there are
two principles related to closed circuits:
1. The volume of the circuit
must be kept constant
2. The oxygen fraction expired
(FEO2) must remain constant, at a value previously
established.
A circuit is considered
“closed” when the volume of oxygen and anaesthetic gas supplied is similar to
the volume uptake by the patient. Therefore, the oxygen supply must equal the
metabolic consumption of the anaesthetized animal and the anaesthetic gases
supply (N2O and halogenated) must equal the uptake in each period of
the procedure.
To corroborate the
impermeability of the circuit, after filling the system with oxygen and closing
both the relief valve and the gas exit (at the “Y” piece), the pressure in the
vacuomanometer must remain at 30 cm of H2O. The loss must remain
under 100 mL/min to be acceptable. However, in small patients or for low volume
circuits, this leak may complicate the use of this technique. Again, these
circuits need methods to supply the anaesthetic vapour independently from the
FGF.
Volumetric addition
The volumetric addition is an
option that demands great attention of the anaesthesiologist, especially during
the first half-hour of the procedure. The injection of the anaesthetic liquid
at previously established times into the inspiratory or expiratory side of the
breathing system will allow maintaining the desired anaesthetic level. The
expiratory side presents some advantages at higher temperatures and a better
dilution of the anaesthetic liquid. Injection times can be easily calculated as
the square of the number of administration. Thus, the Unit Dose (UD in equation
5) should be injected at 0, 1, 4, 9, 16, 25 minutes and so on, which yields 02,
12, 22, 32, 42, 52, etc.
Previous to the injection of the first UD, an initial dose (PD, equation 7)
should be added. Although this method allows keeping an adequate average of the
anaesthetic concentration, there are fluctuations between peaks and valleys
that can alter the anaesthetic depth.
To avoid fluctuations, it may
be more practical to constantly inject the liquid through an infusion pump. In
these cases, the total volume to be administered in the procedure must be
proportionally divided and the pump must be set according to anaesthetic
expired levels (25).
The anaesthetic liquid must be
compatible with the plastic materials of the system. Halogenated compounds
rapidly disintegrate polycarbonate, which is known to be resistant. Other
materials, as polyurethane and nylon are resistant and can be used. With these
techniques, the drug is only used to saturate the peripheral compartments and
keep the FA in the desired level. A thus we observe a significant reduction in
the anaesthetic consumption.
In a horse weighing 550-600
kg, the PD would be about 5 ml and the UD to keep an isoflurane FA similar to
the MAC (1.3 %vol) is about 4 mL (17). In a procedure of 100 minutes, 10 UD (40
mL) would be consumed. Therefore, the total consumption would be about 45 mL of
isoflurane (PD plus 10 UD) in 100 minutes. This volume is much less than the
volume consumed in a system where the vaporizer is outside the circuit that
works with a FGF of 8-10 L/min during the induction phase (about 15-20 min) and
4-5 L/min the rest of the procedure.
In-the-circle
vaporizer (VIC)
The use of closed circuits
with in-the-circle vaporizer is an excellent option that is utilized in
veterinary medicine (26, 27). Stephen’s, Komesaroff’s and Ohio #8, are some of
the vaporizers used with these circuits. These vaporizers are generally located
in the inspiratory side of the circuit. These vaporizers are not precise, have
low resistance, are not temperature compensated and have a glass vaporizer
chamber which allows the alternative use of diverse volatile agents. Although
the anaesthetic concentration in the inspired fraction will depend on many
factors, like room temperature, the ventilation pattern (spontaneous vs. controlled)
and the FGF, the vaporization rate is constant and predictable (28). The
advantages are a decrease in heat and humidity loss, less pollution and a
decrease in anaesthetic consumption. With the exception of desflurane, which
can reach high concentrations at the beginning of the procedure with a flow of
5-10 mL/kg/min.
Stephen’s vaporizer has been
designed for veterinary use and has a dial graduated in eight eighths, which
represent the percentage of the gas column that will pass through the glass
vaporizer chamber to collect the anaesthetic vapour. Usually, the flow draw
over vaporizer chamber is regulated in ¾ ± ¼ during the induction phase and in
3/8 ± 1/8 during the maintenance phase. Fresh gas supply is
about 5-10 mL/kg/min during the first half-hour and between 2.5 and 5 mL/kg/min
for the rest of the procedure. When an in-the-circle vaporizer is used, the
inspired concentration increases with increased temperature and
ventilation (spontaneous or mechanical),
while the Fi decreases if the fresh gas flow increases. Mechanical
ventilation can be utilized with an in-the-circle vaporizer but it is necessary
to strictly control all vital signs.
In small animals (up to 35
kg), the cumulative amount of oxygen and anesthetic used once equilibrium is
obtained is sufficient to last about 20–30 minutes without additional
supplementation. Thus, with a reservoir bag of 3-4 L, the oxygen and
anaesthetic supply can be suspended after 40-60 minutes for an interval of ±20
minutes with only small changes in the anaesthetic depth. The FiO2
is maintained over 45%, which guarantees a correct haemoglobin
saturation. Once the vaporizer is closed, the anaesthetic Fi
decreases very slowly, producing very slow changes in the anaesthetic level.
This process becomes more efficient when the anaesthetic agent has higher blood
solubility and the circuit has greater volume (26).
Accumulation of gases and degradation compounds in closed circuits
The gases that accumulate in
closed circuits can be divided in 4 groups:
1. Compounds formed in the
body: hydrogen, acetone, carbon monoxide (CO) and methane.
2. Compounds absorbed by the
body: ethanol (uncommon in animals), CO and nitrogen.
3. Compounds produced in the
circuit: CO and Component A from halogenated agents.
4. Compounds that enter the
system due to permeability: nitrogen.
From the above mentioned compounds, the most important ones
are those that can alter the patient’s health. The accumulation is not
clinically outstanding in routine practice because the average duration of the
procedure is relatively short. Nevertheless, purging the system can reduce the
concentration of any of the inert gases.
Sevoflurane may react with CO2
absorbents forming the Compound A, which is potentially nephrotoxic and
neurotoxic (31). This compound can accumulate in closed circuits. Therefore,
the use of this agent in procedures involving closed circuits is not
recommended. Although concentrations of the Compound A are generally below the
suggested limits (32), the accumulation of the Compound A can be avoided by
using absorbents like calcium hydroxide, which is free of sodium hydroxide and
potassium hydroxide (33).
Conclusion
The reduction in FGF is
related to a decrease in halogenated agent consumption. Because of this
phenomenon, some compounds which have previously been avoided due to a lack of
economic means can now be integrated into the routine practice. Numerous
studies demonstrate the feasibility of this technique and the few side effects
associated with it.
We consider the low flow
technique to be a valuable therapeutic resource, bringing closer the gap
between modern standards and the daily veterinary practice.
References
1.
Aldrete JA. Closed circuit anaesthesia prevents moderate hypothermia
occurring in patients having extremity surgery. The circular 1987; 4: 3-4.
2.
Kleemann PP. Humidity of anaesthetic gas with respect to low flow
anaesthesia. Anaesth Intens Care 1994; 22: 396-408.
3.
Waters RM. Clinical scope and utility of carbon dioxide filtration in
inhalation anaesthesia. Anesth Analg 1924; 3: 20-28.
4.
Baum Jan. Low Flow Anaesthesia: the theory and practice of low flow,
minimal flow and closed system anaesthesia. 2nd ed. 2001.
Butterworth-Heinemann.
5.
Brody S. Bioenergetics and Growth, New York, 1945.
6.
Lowe H, Ernst EA. The Quantitative Practice of Anaesthesia. Williama
& Wilkins, Baltimore, 1981.
7.
Moens Y. Introduction to the quantitative technique of closed circuit
anaesthesia in dogs. Vet Surg 1988; 17: 98-104.
8.
Olson KN, Klein LV, Nann LE, Soma LR: Closed-circuit liquid injection
isoflurane anaesthesia en the horse. Vet Surg 1993; 22: 73-78.
9.
Harper NJN: Is oxygen an acceptable carrier gas? Symposium, Association
for Low Flow Anaesthesia, Pisa, Italy, April 26 - 27, 2002.
10.
Akca O, Podolsky A, Eisenhuber E, Panzer O, Hetz H, Lampl K, Lackner
FX, Wittmann K, Grabenwoeger F, Kurz A, Schultz AM, Negishi C, Sessler DI..
Comparable postoperative pulmonary atelectasis in patients given 30% or 80%
oxigen during 2 hours after colon resection. Anaesthesiology 1999; 91: 991-998.
11.
Greif R, Akca O, Horn EP, Kurz A, Sessler DI. Supplemental
perioperative oxygen to reduce the incidence of surgical-wound infection.
Outcomes Research Group. N Engl J Med 2000; 342:161-7.
12.
Kotani N, Hashimoto H, Sessler DI, Muraoka M, Hashiba E, Kubota T,
Matsuki A. Supplemental intraoperative oxygen augments antimicrobial and
proinflammatory responses of alveolar macrophages. Anaesthesiology 2000; 93:
15-25.
13.
Barber RE, Lee J, Hamilton WK. Oxygen toxicity in man. A prospective
study in patients with irreversible brain damage. New Eng J Med 1970; 283:
1478.
14.
Lowe H. The anaesthetic continuum. In Aldrete JA, Lowe HJ, Virtue RW,
eds. Low flow and closed system anaesthesia. Grune & Stratton, New York,
1979.
15.
Eger EI, II. Anaesthetic Uptake and Action. Baltimor. Williams &
Wilkins, 1974.
16.
Aldrete JA, Romo Salas F. Oxigenation with high, intermediate and low
gas flows during toraxic and abdominal surgery. Studie at the altitude of one
mile in low flow and closed system anaesthesia. In: Aldrete JA, Lowe HJ, Virtue
RW, eds. Low flow and closed system anaesthesia. Grune & Stratton, New
York, 1979.
17.
Foldes FF, Ceravolo AJ, Carpenter SL. The administration of nitrous
oxide-oxigen anaesthesia. Ann Sur 1952; 136:978-981.
18.
Virtue RW. Minimal flow ntrous oxide anaesthesia. Anestesiology
1974; 40: 196-198
19.
Aldrete JA. Fron the theory to the practice of “near-closed
system”anesthesia. Appl Cardiop Pathophy 5 Suppl. 1995; 2:5-11.
20.
Muir WW III, Hubbell JAE. Handbook of veterinary anaesthesia. St Louis:
CV Mosby, 1989.
21.
Wagner AE, Bednarski RM. Use of low flow and closed-system anaesthesia.
JAVMA 1992; 200: 1005-1011.
22.
Muir WW III, Hubbell JAE. Handbook of veterinary anaesthesia. 2da
edición. St Louis: CV Mosby, 1989.
23.
D’Avirro S. Anestesia inhalatoria con circuito cerrado y flujos
mínimos. Rev Arg Anest 2000; 58: 163-173.
24.
Aldrich J, Haskins S. Monitoring the critically ill patient. In Kirk
XII, "Current Veterinery Practice" Saunders 1995.
25.
Nunn G. Flow independent delivery of volatile agents. Symposium,
Association for Low Flow Anaesthesia, Pisa, Italy, April 26 - 27, 2002.
26.
Haskins S.C.(1992a) The Case for In Circuit Vaporisers. Opinions in
Small Animal Anaesthesia. Veterinary Clinics of North America. 1992; 22:
391-392.
27.
Otero P, Guerrero J, Bonafine R, Hallu R: Evaluacion de un sistema
anestésico cerrado con vaporizador
dentro del circuito, en caninos anestesiados
con halotano e isofluorano. Selecciones Veterinarias 2001; 9:36-42.
28.
Bednarski RM, Gaynor JS, Muir III WW: Vaporiser en circle for delivery
of isofluorane to dogs. J Am Vet Med Assoc 1993; 202:943-948.
29.
Laredo FG, Cantalapiedra AG, Agut A, Pereira JL, Murciano J: The
Komerasoff anaesthetis machine for delivering sevoflurane to dogs. Vet Anaesth
Analg 2001; 28:161-167.
30.
Otero, P, Pisera, D, Hallu, R, Aldrete JA. Sevofluorane anaesthesia in dogs under
closed circuit system. Poster presentado en: Symposium, Association for Low
Flow Anaesthesia, Pisa, Italy, April 26 - 27, 2002.
31.
Gentz BA, Malan TO Jr. Renal toxicity with sevoflurane: a storm in a
teacup? Drugs 2001; 61: 2155-62.
32.
Muir III WW, Gadawski J: Cardiorespiratory effects of low-flow and
closed circuit inhalation anaesthesia, using sevoflurane delivered with an
in-circuit vaporiser and concentrations of compound A. Am J Vet Res 1998;
59:603-608.
33.
Versichelen LF, Bouche ML, Rolly G, Bocxlaer JF, Struys MM, De Leenheer
AP, Mortier EP: Only carbon dioxide absorbents free of both NaOH and KOH do not
generate compound A in vitro closed-system sevoflurane. Anaesthesiology 2001;
95:750-5.
34.
Eger EI, II. Complexities overlooked: things may not be what they seem.
Anesth Analg 1997; 239-240.
35.
Boller M, Moens Y, Kaestner S, Bettschart-Wolfensberger R. Closed
system anaesthesia in dogs using liquid injection of sevoflurane: evaluation of
the square-root-of-time model and the influence of the CO2
absorbent. Congress Proceedings, 8th World congress of veterinary
anesthesia. Knoxille, Tennessee, USA. 2003: 189.